DO YOU KNOW...

CADD® Solis pump alarms

 

The following will help you know what to do when the CADD Solis pump alarms.

CADD Solis pump alarm blue
CADD Solis pump alarm red
CADD Solis pump alarm

Blank screen (High)

Batteries removed or door opened.

  1. Turn pump on.
  2. Clear alarms if needed.

Air in line (High)

  1. Select Acknowledge.
  2. Close clamps.
  3. Disconnect line from patient.
  4. Prime tubing.
  5. Clean patient’s line.
  6. Reconnect tubing.
  7. Press the stop/start key.
CADD Solis pump alarm

Battery depleted (High) or battery low (Low)

  1. Select Acknowledge.
  2. Press start/stop key to stop the pump.
  3. Install 4 new AA batteries.
  4. Press the power switch.
  5. Press the stop/start key.
  6. Reset reservoir volume. (yes/no)
  7. Start pump? (yes)
  8. Wait for pump to start.

Downstream occlusion (blockage) (High) or upstream occlusion (High)

  1. Press the stop/start key to silence.
  2. Check for blockages, kinks, and closed or capped tubing.
  3. Press the stop/start key to restart pump.

Reservoir volume low (Medium/Low)

  1. Select Acknowledge.
  2. Prepare to install a new bag.
CADD Solis pump alarm

Reservoir volume zero (High)

  1. Select Acknowledge.
  2. Prepare to install a new bag.
  3. Follow these steps to attach new bag.
    • Press the stop/start key to stop the pump.
    • Clamp tubing.
    • Push down on lever to detach latch.
    • Remove and discard cassette, tubing, and bag.
    • Attach new cassette, tubing, and bag.
      • Remove blue clip.
      • Insert hooks into hinge pins and secure cassette.
      • Lift up on the lever to the closed position.
      • Pump will display directions.
  4. Press stop/start key to restart pump
  5. Reset reservoir volume. (yes/no)
  6. Start pump? (yes)
  7. Wait for pump to start.

Questions?

If you have questions or concerns while using the CADD Solis pump, contact your child’s home infusion nurse.


 

This document is not intended to take the place of the care and attention of your personal physician or other professional medical services. Our aim is to promote active participation in your care and treatment by providing information and education. Questions about individual health concerns or specific treatment options should be discussed with your physician.

St. Jude complies with health care-related federal civil rights laws and does not discriminate on the basis of race, color, national origin, age, disability, or sex.

ATTENTION: If you speak another language, assistance services, free of charge, are available to you. Call 1-866-278-5833 (TTY: 1-901-595-1040).

ATENCIÓN: si habla español, tiene a su disposición servicios gratuitos de asistencia lingüística. Llame al 1-866-278-5833 (TTY: 1-901-595-1040).

تنبيه: إذا كنت تتحدث باللغة العربية فيمكنك الاستعانة بخدمات المساعدة اللغوية المتوفرة لك مجانا. .يرجى الاتصال بالرقم. 5833-278-866-1  (الهاتف النصي: 1040-595-901-1).